Introduction of OIL-8 Detection Oil Content in Water Detector
Introduction of OIL-8 Detection Oil Content in Water Detector
When a mono chromatic light through the test solution,the light will be absorbed, light be absorbed proportional to the concentration of the test solution.
Conform Beer's law:
A=log(1/T)=log(I0/I)
T:Transmission
I0: Incident light intensity
I: Transmitted light intensity
A: Absorbance
Oil concentration:
C=X×A2930+Y×A2960+Z×(A3030-A2930/F)
C: Oil concentration
A2930A2960A3030: Absorbance under different wavelengths
X,Y,Z,F: Correction factor
Main specifications of OIL-8 Detection Oil Content in Water Detector
Measurement range
|
0.12 mg/L~640mg/L ( 5cm~0.5cm Cuvette)
|
Detection limit
|
<0.12mg/L(CCl4 Extract,using 5cm cuvette testing directly
|
<0.002mg/L(Water sample,Extraction ratio 100:1, 5cm cuvette)
|
Max. Measurement Concentration
|
64000mg/L
|
Measuring accuracy
|
±0.5 (C<10 mg/L)
|
±2% (10≤C≤80mg/L)
|
Zero stability
|
±1% 24 hours
|
Stability
|
≤1% (20mg/L oil sample)
|
Linear
|
R>0.999
|
Wave number range
|
3400cm-1~2400cm-1
|
Absorbance range
|
0.0000~3.000AU
|
Baseline drift
|
Absorbance <0.001AU/60min
|
Component
|
optical system, electrical system, micro-computer and data processing working station
|
Overall Size
|
550mm*380mm*150mm
|
Weight
|
15Kg
|
Power
|
(220±22)V/(50±1)Hz/35VA
|
Using environment
|
5ºC~35ºC 20%~80%RH
|
Photos of OIL-8 Detection Oil Content in Water Detector
因产品生产批次、具体型号不同,以上图片仅供参考,详情可联系我们的销售人员进行具体核实。